100 Hz Noise VEP Signal on CRT

100 Hz noise on a VEP result displayed on a CRT monitor. 

 

Explanation: Monitor cross - talk

 

Action: Move patient further away from monitor. Measure the distance between patient - monitor >

from test window> Input distance into Steps tab > Stimulator > Distance. 

from protocol parameters > Input distance into Stimulus tab > View distance. 